Infrared detector without diffracting element dispersing the IR wave lengths
A nondispersive infrared sensor (or NDIR sensor) is a simple
spectroscopicsensor often used as a
gas detector. It is non-dispersive in the fact that no dispersive element (e.g a
prism or
diffraction grating as is often present in other
spectrometers) is used to separate out (like a
monochromator) the broadband light into a narrow spectrum suitable for gas sensing. The majority of NDIR sensors use a broadband lamp source and an optical filter to select a narrow band spectral region that overlaps with the absorption region of the gas of interest. In this context narrow may be 50-300nm bandwidth. Modern NDIR sensors may use
Microelectromechanical systems (MEMs) or mid IR
LED sources, with or without an
optical filter.
Principle
The main components of an NDIR sensor are an
infrared (IR) source (lamp), a sample chamber or
light tube, a
light filter and an
infrared detector. The IR light is directed through the sample chamber towards the detector. In parallel there is another chamber with an enclosed reference gas, typically
nitrogen. The gas in the sample chamber causes
absorption of specific wavelengths according to the
Beer–Lambert law, and the attenuation of these wavelengths is measured by the detector to determine the gas concentration. The detector has an
optical filter in front of it that eliminates all light except the wavelength that the selected gas molecules can absorb.
Ideally other gas molecules do not absorb light at this wavelength, and do not affect the amount of light reaching the detector however some cross-sensitivity is inevitable.[1] For instance, many measurements in the IR area are cross sensitive to
H2O so gases like
CO2,
SO2 and
NO2 often initiate cross sensitivity in low concentrations.[citation needed][2]
The IR
signal from the source is usually chopped or modulated so that thermal background signals can be offset from the desired signal.[3]
Configurations with multiple filters, either on individual sensors or on a rotating wheel, allow simultaneous measurement at several chosen wavelengths.
Fourier transform infrared spectroscopy (FTIR), a more complex technology, scans a wide part of the spectrum, measuring many absorbing species simultaneously.
Research
Miniature IR sources based on
microelectromechanical systems (MEMS) have been experimentally applied to NDIR systems since 2006 and is useful since 2016. The low energy of MEMS emission means a sensitive detector circuit based on lock-in amplification is needed.[4] Other useful detectors include the
photoacoustic gas sensor which use a MEMS microphone to detect IR-gas interactions.[5]
Gases and their sensing wavelengths
Gases do not have a specific sensing wavelength, rather there are regions of the IR spectrum where there are typically many thousands of closely spaced absorption lines. See the
Hitran database for more information.
NO — 5.3 μm, NO2 has to be reduced to NO and then they are measured together as NO x; NO also absorbs in ultraviolet at 195-230 nm, NO2 is measured at 350-450 nm;[8] in situations where NO2 content is known to be low, it is often ignored and only NO is measured; also, 1.8 μm[6]
^"2.2 Nitrogen oxide (NOx)". 2. Measuring instruments for stationary source, Air Pollution Continuous Monitoring Technology in Japan. Global Environment Centre Foundation. Archived from
the original on 2017-09-16. Retrieved 2020-01-16.